Performance
Tests
C-10
Procedure Steps
Step
1.
Check
Deflection
Accuracy
a.
Connect
a 10
mV
standard-amplitude
signal
from
the
calibration
generator
via
a 50 Q BNC
coaxial
cable
to
the
CH 1 OR X
input
connector.
b. CHECK -
Deflection
accuracy
is
within
the
limits
given
in
Table
C-2 for
each
CH 1 VOL TS/DIV
switch
setting
and
corresponding
standard-amplitude
signal.
Step
2.
Check
Trace
Shift
a.
Set
CH 1 VOLTS/DIV
switch
to 2
mV/division.
b. Rotate
the
VOLTS/DIV
switch
through
the
range.
c. CHECK -
The
trace
shift
should
be
$ 1.0
division.
d.
Set
the
VAR
to
the
minumum
sensitivity,
and
rotate
the
VOLTS/DIV
switch
through
the
range.
e.
CHECK -
The
trace
shift
should
be$
1.0
division.
f.
Move
the
cable
from
the
CH 1 OR X
input
connec-
tor
to
the
CH 2
ORY
input
connector.
Toggle
the
MODE
switch
to CH
2.
g.
Repeat
part
b
through
fusing
the
CH 2
controls.
h.
Push
the
CH 2
INVert
switch
to
the
INV
position.
j. CHECK -
The
trace
shift
should
be$
1.5
division.
Rev
12/92
Appendix
C:
Performance Verification

Table of Contents

Other manuals for Tektronix 2212

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2212 and is the answer not in the manual?

Tektronix 2212 Specifications

General IconGeneral
BrandTektronix
Model2212
CategoryTest Equipment
LanguageEnglish

Related product manuals