i u i a   ^ .  r c i  l u m t a i i w   » « i  u i u m t o u
P I  Step 2:  Check Store Differentia! and Cursor Time Difference Accuracy
a.
b.
c.
d.
e.
f.
g-
h.
i.
k.
m.
n.
o.
p.
Set: CH 1  AC-GND-DC 
SEC/DIV 
X I0 Magnifier 
STORE/NON-STORE
GND
0.1  ms
Off (knob in)
STORE (button in)
Use the CH 1  POSITION control to center the base line vertically 
and the Horizontal POSITION control to align the start of the trace 
with the 1st vertical graticule line.
Using the CURSORS control and SELECT C1/C2 (push in the 
CURSORS control knob) switch, select one of the two cursors and 
set it exactly on the 2nd vertical graticule line. Select the other cursor 
and move it towards the right until the AT readout displays 
0.800 ms,
CHECK— Graticule indication of cursor difference at the 10th verti
cal graticule line is within 0.16 division.
Set the CH 1  AC-GND-DC switch to DC.
Select 0.1  ms time markers from the time-mark generator.
Align the 2nd time marker with the 2nd vertical graticule line using the 
Horizontal POSITION control.
Press in the SAVE/CONT button to select SAVE for a stable display.
Use the CURSORS control and SELECT C1/C2 (push in the CUR
SORS control knob) switch to set the first cursor on the trailing edge 
of the 2nd time marker.
Press in the CURSORS control knob again to activate the second 
cursor.
Set the second cursor on the trailing edge of the 10th time marker at 
the same voltage level as on the 2nd time marker.
CHECK — The AT readout is between 0.798 ms and 0.802 ms.
Press in the SAVE/CONT button to select CONT.
Set the SEC/DIV switch to 0.5 ps.
Select 0.5 ps time markers from the time-mark generator.
Align the 2nd time marker with the 2nd vertical graticule line using the 
Horizontal POSITION control.
NOTE
Allow the points to accumulate for a few seconds before saving the 
display.
q.  Repeat parts h through k.
►
f
*
I
I
i
Z
I
£
F
fe.
I
F
ttt
£
it
K
h
&
I
I
I
I
I
A-40
Appendices
t