Appendix C: Performance Verification
NOTE
Pulses with fast rise and fall times have only a few sample points 
and it may not be possible to place the cursors at exactly the same 
voltage levels.
r.  CHECK — The AT readout is between 3.990 ps and 4.010 ps. 
Q   Step 3:  Check Variable Range
b.  Select 0.5 ms time markers from the time-mark generator.
c.  CHECK — Time markers are one division or less apart.
FH  Step 4:  Check Position Range
b.  Select 10 ps time markers from the time-mark generator.
c.  CHECK — Start of the sweep can be positioned to the right of the 
center vertical graticule line by rotating the Horizontal POSITION 
control fully clockwise.
d.  CHECK — The 11th time marker can be positioned to the left of the 
center vertical graticule line by rotating the Horizontal POSITION 
controi fully counterclockwise.
e.  Select 50 ps time markers from the time-mark generator.
f.  Align the 3rd time marker with the center vertical graticule line using 
the Horizontal POSITION control.
g.  Set the X10 Magnifier knob to On (knob out).
h.  CHECK — Magnified time marker can be positioned to the left of the 
center vertical graticule line by rotating the Horizontal POSITION 
control fully counterclockwise.
i.  CHECK — Start of the sweep can be positioned to the right of the 
center vertical graticule line by rotating the Horizontal POSITION 
control fully clockwise.
n   Step 5;  Check Store Expansion Range
a.  Set:
SEC/DIV 
SEC/DIV Variable 
STORE/NON-STORE
0.2 ms
Fully counterclockwise 
NON-STORE (button 
out)
a.  Set; 
SEC/DIV
10 ps
a.  Set:
SEC/DIV 
X I0 Magnifier
0.1  ms 
Off (knob in)
I
2221A User Manual
A-41