Appendix C: Performance Verification
b.  Select 10 ps time markers from the time-mark generator.
c.  Use the Horizontal POSITION control to align the start of the sweep 
with the 1st vertical graticule line.
d.  Set the STORE/NON-STORE switch to STORE (button in).
e.  Set the X I0 Magnifier knob to On (knob out).
f.  CHECK — The time markers are one division apart.
□   steP 6:  Check 4K to 1K Display Compress
a.  Set:
SEC/DIV  50 ps
X I0 Magnifier  Off (knob  in)
1K/4K  4K
b.  Select 0.1  ms time markers from the time-mark generator and check 
that the time markers are two divisions apart,
c.  Rotate the SEC/DIV Variable control out of detent.
d.  CHECK — For two time markers per division over the center eight 
divisions.
n   Step 7:  Check Non-Store Sweep Length
a.  Set:
SEC/DIV Variable  CAL detent
STORE/NON-STORE  NON-STORE (button
out).
b.  Use the Horizontal POSITION control to align the start of the sweep 
with the 1st vertical graticule line.
c.  CHECK — End of the sweep is to the right of the 11 th vertical grati
cule line.
d.  Disconnect the test equipment from the instrument.
F I  Step 8:  Check X Gain
a.  Set:
X-Y  On (button in)
CH 1 VOLTS/DIV  10 mV
Horizontal POSITION  Midrange
b.  Connect the standard-amplitude signal from the Calibration Genera
tor via a 50 Q cable to the CHI  OR X input connector.
c.  Set the generator to produce a 50 mV signal.
d.  Use the CH 2 POSITION and Horizontal POSITION controls to 
center the display.
e.  CHECK — Display is 4.85 to 5.15 horizontal divisions.