Appendix
B:
Specification
Table A-4: Electrical Characteristics (Cont.)
Characteristics Performance Requirements
Aberrations (NON-STORE and STORE in De-
fault Modes)
2
mV per division to 50 mV per division
+4%, -4%, 4% p-p
0.1 V per division to 0.2 V per division
+6%, -6%, 6% p-p
0.5 V per division
1 V per division to
5
V per division +12%, -12%, 12% p-p1
Measured with a vertically centered six-division reference
signal, from a 50
Q
source. The source is connected through
a 50
Q
coaxial cable terminated in 50
Q
at the input connec-
tor. The
VOLTSIDIV
Variable control is in the
CAL
detent.
Useful Storage performance2
RECORD, SCAN and ROLL Store Modes
SAMPLE Acquisition, no AVERAGE
1
ps per division to
5 s per division
Single Trace
10
Hz1
SECIDIV setting
5
Hz1
SECIDIV setting
EXT CLOCK (up to 100 kHz)
PEAK DETECT
Sine Wave 10
MHZ'
Amplitude Capture
(5%
p-p
maximum amplitude uncertainty)
Pulse Width 10 ns
Amplitude Capture (50% p-p
maximum amplitude uncertainty)
1
Performance requirement not checked in manual.
2Useful storage performance is limited to the frequency where there are
10
samples per sine wave signal period at the maximum
sampling rate. (Maximum sampling rate is
100
MHz.) This yields a maximum amplitude uncertainty of
5%.
Accuracy at the useful
storage bandwidth limit is measured with respect to a six-division,
50
kHz reference sine wave.
A-70
Appendices