EasyManuals Logo

Tektronix Keithley 4225-PMU User Manual

Tektronix Keithley 4225-PMU
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Model 4200A
-SCS Pulse Card (PGU and PMU) User's Manual Section 2:
Connections
4200A-PMU-900-01 Rev. B March 2023 2-7
The RPM is matched to a PMU card and channel. Make sure to connect the RPM only to that PMU
card and channel. Labels on top of the RPM indicate the specific connection channel and PMU. For
example, "PMU1-2" indicates PMU1, channel 2.
With system power off, use the supplied RPM cable to connect a 4225-RPM to the matching RPM
channel of the 4225-PMU, as shown in the following figure.
Figure 6: 4225-PMU connection to an RPM
With an RPM installed, never make connections directly to any of the SMA connectors (CH1
and CH2) on the PMU. This may result in damage to the PMU or DUT. It may also produce
corrupt data.
After connecting or removing an RPM, always perform the procedure “Update the RPM
configuration” in Model 4200A-SCS Setup and Maintenance to ensure that KCon accurately
represents the present 4200A-SCS hardware configuration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4225-PMU and is the answer not in the manual?

Tektronix Keithley 4225-PMU Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4225-PMU
CategoryComputer Hardware
LanguageEnglish