EasyManua.ls Logo

Tektronix TDS5000B Series

Tektronix TDS5000B Series
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands
5-8
TDS5000B Series Service Manual
Table 5- 4: CALibrate commands (Cont.)
Command Description
CALibrate:INTERNal:STATus? Returns t he current status of the internal signal path
calibration
CALibrate:PROBEstate:CH<x> Returns t he cali bration st atus of the attached probe
CALibrate:RESults? Returns t he status of all calibration subsystems
without performing an SPC operation
CALibrate:RESults SPC? Returns t he results of the last SPC operation
*CAL?
This query-only command instructs the digitizing oscilloscope to perform a
signal path compensation and return its calibration status. This command is
equivalent to selecting Instrument Calibration from the Utilities menu and then
clicking the C alibrate button.
NOTE. The self-calibration can take several minutes to r espond. No other
commands will be executed until calibration is complete.
CALibrate:RESults:SPC?
*CAL?
0 <NR1> indicating that the calibration did not complete successfully
1 (or any non-zero value) <NR1> indicating that the calibration completed
without errors
*CAL?
might return
1
indicating that the calibration was successful.
Related C ommands
Syntax
Returns
Example

Table of Contents

Other manuals for Tektronix TDS5000B Series

Related product manuals