Commands
TDS5000B Series Service Manual
5-25
<NR1> sets the index item about which data will be returned, which ranges from
0 through 15.
DIAg:ITEM:SUBITEMS? 2
This query might return
:DIAG:ITEM:SUBITEMS 2,8
This indicates that there are 8 sub-items associated with the selected item (2).
DIAg:LEVEL
This command sets or returns the currently selected level of diagnostic test
hierarchy. This command is equivalent to selecting Instrument Diagnostics from
the Utilities menu and then reviewing the Diagnostic Status.
DIAg:LEVEL (SUBSYS|AREA|TEST)
DIAg:LEVEL?
This query form of the command can be issued while diagnostics are still in
progress.
SUBSYS sets diagnostic testing to the subsystem level.
AREA sets diagnostic testing to the area level.
TEST sets diagnostic testing to the test level.
DIAg:LEVEL AREA
sets the diagnostic testing to the area level.
DIAg:LEVEL?
This query might return
:DIAG:LEVEL SUBSYS
indicating that the current level of diagnostic test hierarchy is Subsys.
Arguments
Example
Syntax 1
Syntax 2
Arguments
Example 1
Example 2

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals