Commands
5-26
TDS5000B Series Service Manual
DIAg:LOOPS?
This query-only command returns the number of times that the selected
diagnostics set was completed during the last diagnostic execution. This
command is equivalent to selecting Instrument Diagnostics from the Utilities
menu and then reviewing the Elapsed Count.
This command can be issued while diagnostics are still in progress.
DIAg:LOOPS?
DIAg:LOOPS?
This query might return
:DIAG:LOOPS 5
indicating that the selected set of diagnostics was completed 5 times during the
last diagnostic execution.
DIAg:NAMe?
This query-only command returns the names of the subsystem, area and test of
the current diagnostic test.
This query can be issued while diagnostics are still in progress.
DIAg:NAMe:AREA?, DIAg:NAMe:SUBSYS?, DIAg:NAMe:TEST?
DIAg:NAMe?
DIAg:NAMe:?
This query might return
:DIAG:NAME:SUBSYS Processor;AREA Memory;TEST DRAMWalk1
This indicates the subsystem name, area name, and test name of the currently
selected diagnostic test.
Syntax
Example
Related C ommands
Syntax
Example

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals