Removal and Installation Procedures
TDS5000B Series Service Manual
6--15
1. Remove the right side cover: See Figure 6--4.
H Trim (all)
H Bottom cover
2. Orient the oscilloscope: Set the oscilloscope so the rear is on the work
surface and the front of the oscilloscope is facing you.
NOTE. All mounting screw holes are indicated by a star etched around the
mounting hole except two that are marked with two tick marks. Be sure to
properly locate these screws to reinstall the right side cover.
a. Remove the fourteen T-15 Torxdrive screws that secure the right-side
cover to the top and right sides of the chassis.
b. Remove the right-side cover.
CAUTION. Take care not to bind or snag the covers on the oscilloscope internal
cabling as you remove or install.
CAUTION. To prevent damage to your instrument, be sure to properly locate the
black screw to reinstall the right-side cover.
3. Reinstallation: Do in reverse steps a through b to reinstall the right-side
cover. Be sure to properly locate the black screw (if present; see Figure 6--4).
Right Side Cover