EasyManua.ls Logo

Ulvac GI-M2 - Selecting Thes; Unction; Setting the Sensitivity Function

Default Icon
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
8.4.Selecting the Sensitivity Factor Function
Refer to P.35
Set a sensitivity factor. The set point as follows.
GI-M2
GI-D7
GI-N8
SEn:1
N2
N2
N2
SEn:2
Ar
Ar
H2
SEn:3
Any value
Any value
Any value
SENS-①
The portion in the left figure blinks
after the segment indicator displays [
SEn
__
] for about 1 second. Each time the []
key is pressed, the portion in the left
figure repeats displaying [
SEn _1
]
[
SEn _2
] [
SEn _3
] [
SEn _1
] ...
cyclically.
To use the sensitivity factor for N2, press
the [(ENTER)] key, with [
SEn _1
] [
SE
_2
] selected, to transfer control to the
SET1-
status.
To use the sensitivity factor for OTHER,
press the [(ENTER)] key, with [
SEn _3
]
selected, to transfer control to the SET1-
status.
CAUTION:
When setting a sensitivity factor, the set point action may change over depending
on the set value the moment the N2 setting is changed over to Ar setting or OTHER
setting. It is recommended to change the setting in the POW-
status or by
stopping components from which signal is received. Check the set point set value
and sensitivity factor set value before changing the setting.
8.5.Setting the Sensitivity Factor Function
Refer to P.35
Setting the sensitivity factor function. The set point as follows.
GI-M2
GI-D7
GI-N8
Setting
range
6.00×10
-1
6.00×10
-3
7.47×10
-1
1.66×10
-2
2.00×10
-0
.00×10
-2
SENS-②
The
portion in the left figure blinks.
Each press on the [
] key changes the
numeric value of the blinking portion.
When a desired value is displayed,
press the [
(ENTER)] key to transfer
control to the SEN-
status.

Table of Contents

Other manuals for Ulvac GI-M2