11.4. Explanation of Command Action
11.4.1. Changing over remote/local control
Command
Parameter
Response
Explanation of function
‘RE’
‘OK’ or ‘NG’
Sets remote status
‘LO’
‘OK’ or ‘NG’
Sets local status
Send RE command when the remote control is used in the remote (RS-232C) mode.
After recept RE, lit when in the LED of “REMOTEon a front panel and send back OK’.
When the remote control by the communication, a front panel and the input from IO are
invalid.
Send LOwhen it finish the remote control by communications.
Use the following control commands after the ,`’RE ' sending.
11.4.2. Reading the connection model
Command
Parameter
Response
Explanation of function
‘GS’
‘GI-
xx
Reads control status
Send back the model (the model name).
Send back ‘GI-M2 GI-D7
11.4.3. Control of filament
Refer to P.40
Command
Parameter
Response
Explanation of function
‘FO’
‘OK’ or ‘NG’
Turns OFF filament
‘F1’
‘OK’ or ‘NG’
Turns ON filament
‘FA’
‘OK’ or ‘NG’
Selects filament-1
‘FB’
‘OK’ or ‘NG’
Selects filament-2
Turning ON/OFF the filament.
When it dose not appoint filament 1/2, it turns on a set filament last.
It is memorized, and the setting of filament 1/2 is reflected at the time of the front panel
operation by the main body.
By a `RS' command, the reading of the set point of the filament is possible.
11.4.4. Control of degassing
Refer to P.31
Command
Paramete
r
Response
Explanation of function
‘D0’
‘OK’ or ‘NG’
Turns OFF degassing
‘D1’
‘OK’ or ‘NG’
Turns ON degassing
‘DA’
‘OK’ or ‘NG’
Selects degassing AUTO mode use ON
‘DB’
‘OK’ or ‘NG’
Selects degassing AUTO mode use OFF
‘DS’
XX
‘OK’ or ‘NG’
Sets degassing AUTO mode time in XX minutes
Degassing is active after the filament is ON.
Setting of AUTOMODE time is 01~99min. Send DS
XX
XX
:01-99 command.
When it dose not set, it uses the value that is finished with the setting of the main body.
It is memorized, and this setting is reflected at the time of the front panel operation by the
main body.