67
11.4. Explanation of Command Action
11.4.1. Changing over remote/local control
-
-
Send ′RE′ command when the remote control is used in the remote (RS-232C) mode.
After recept ‘RE’, lit when in the LED of “REMOTE” on a front panel and send back ‘OK’.
When the remote control by the communication, a front panel and the input from IO are
invalid.
Send ‘LO’ when it finish the remote control by communications.
Use the following control commands after the ,`’RE ' sending.
11.4.2. Reading the connection model
-
Send back the model (the model name).
Send back ‘GI-M2’、 ‘GI-D7’
11.4.3. Control of filament
→ Refer to P.40
-
-
-
-
Turning ON/OFF the filament.
When it dose not appoint filament 1/2, it turns on a set filament last.
It is memorized, and the setting of filament 1/2 is reflected at the time of the front panel
operation by the main body.
By a `RS' command, the reading of the set point of the filament is possible.
11.4.4. Control of degassing
→Refer to P.31
-
-
-
Selects degassing AUTO mode use ON
-
Selects degassing AUTO mode use OFF
Sets degassing AUTO mode time in ′XX′ minutes
Degassing is active after the filament is ON.
Setting of AUTO・MODE time is 01~99min. Send ‘DS
XX
’(
XX
:01-99) command.
When it dose not set, it uses the value that is finished with the setting of the main body.
It is memorized, and this setting is reflected at the time of the front panel operation by the
main body.