EasyManua.ls Logo

Ulvac GI-M2 - Selection of Analog Unction; Setting of Analogf

Default Icon
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
8.12.Selection of Analog Function
Refer to P.41
Setting of analog function. The set point as follows.
GI-M2
GI-D7
GI-N8
rEC:1
Eeach-digit linear output mode
rEC:2
Pseudo-logarithmic mode
rEC:3
Range hold liniea mode
rEC:4
logarithmic mode
GI-D6 compatibility output mode
Not used
rEC
5
Not used
GI-TL3 compatibility output mode
Not used
REC-①
The
portion in the left figure blinks
after the segment indicator displays [
rEC
__
] for about 1 second. Each time the []
key is pressed, the
portion of the
segment indicator in the left figure displays
[
rEC _1
][
rEC _2
][
rEC _
][
rEC _1
]・・ cyclically.
To use recorder LIN(LOG), press the [
(ENTER)] key, with [
rEC _1
]
selected, to transfer control to the BPS-
status.
To use analog REC-HOLD, press the [
(ENTER)] key, with [
rEC _3
]
selected, to transfer control to the REC-
status.
8.13. Setting of Analog Function
Refer to P.41
Set the REC-HOLD numeric values. The set point as follows.
GI-M2
GI-D7
GI-N8
Setting
range
-0
-8
-1
-7
-1
-9
Settable numeric values are sensitivity factor N2. This setting range will change if a
sensitivity factor other than N2 is designated or if sensitivity setting is entered.
REC-②
Each time the [] key is pressed, the
numeric value of the blinking portion
changes. When a desired value is
displayed, press the [(ENTER)] key to
transfer control to the BPS- status.

Table of Contents

Other manuals for Ulvac GI-M2