11.3. List of Commands
Comm
and
Parameter
Response
Explanation of function
‘RE’
‘OK’ or ‘NG’
Sets remote status
LO’
‘OK’ or ‘NG’
Sets local status
‘GS’
‘GI-
xx
Reads control status
FO’
‘OK’ or ‘NG’
Turns OFF filament
‘F1’
‘OK’ or ‘NG’
Turns ON filament
‘FA’
‘OK’ or ‘NG’
Selects filament - 1
‘FB’
OK’ or NG
Selects filament - 2
‘D0’
‘OK’ or ‘NG’
Turns OFF degassing
‘D1’
‘OK’ or ‘NG’
Turns ON degassing
DA’
‘OK’ or ‘NG’
Selects degassing AUTO mode use ON
DB’
‘OK’ or ‘NG’
Selects degassing AUTO mode use OFF
‘DS’
XX
‘OK’ or ‘NG’
Sets degassing AUTO mode time in
XX
minutes
'
EM
'
'
OK
' or '
NG
'
Reads emission valid status
'
ES
'
'
05
' or '
50
'
Reads emission status (GI-N8)
'
E0
'
'
OK
' or '
NG
'
Selects emission current 0.5mA (GI-N8)
'
E1
'
'
OK
' or '
NG
'
Selects emission current 0.5mA (GI-N8)
‘SA’
‘OK’ or ‘NG’
Sets sensitivity factor (N
2
)
‘SB’
‘OK’ or ‘NG’
Sets sensitivity factor (Ar)
‘SE’
X.XX
E
±
XX
’’
‘OK’ or ‘NG’
Sets sensitivity factor
‘R1’
X.XX
E-
XX
’’
Reads data of setpoint-1
‘R2’
X.XX
E-
XX
’’
Reads data of setpoint-2
‘S1’
X.XX
E-
XX
’’
‘OK’ or ‘NG’
Sets setpoint-1
‘S2’
X.XX
E-
XX
’’
‘OK’ or ‘NG’
Sets setpoint-2
‘SP
‘1-
X
/2-
X
Reads setpoint-1 and -2 status
‘LI’
‘OK’ or ‘NG’
Selects analog LIN
LG’
‘OK’ or ‘NG’
Selects analog Pseudo-LOG
LH’
‘OK’ or ‘NG’
Selects analog REC-HOLD
‘L4’
‘OK’ or ‘NG’
Selects Analog LOG output(GI-M2)
‘L6’
‘OK’ or ‘NG’
Selects Analog GI-D6 compatibility output(GI-D7)
‘L3’
‘OK’ or ‘NG’
Selects Analog GI-TL3 compatibility output(GI-D7)
‘LS’
‘E-
XX
‘OK’ or ‘NG’
Sets range hold linear output value (range)
‘RP
‘X.XXE-XX’
Reads measured pressure data
‘PR’
‘ONor ‘OF
Reads EXT-PROTECT status
‘RS
XXXXXXX
Reads action status