9.6.5. Range Hold Linear Output (REC-HOLD)
As in the LIN mode, the mantissa part 3-digits display A. BC and 2-digits display A. BG
are outputted as A. BC [V] and A. BG [V] respectively, but above the specified range of
REC-HOLD, it is outputted as FULL (10 V) and, below the specified range, the action is
the same as when range hold is activated in the LIN range.
(Example) When REC-*HOLD by 10
-4
A.BC×10
-3
10.00(V)
A.BC×10
-4
A.BC(V)
A.BC×10
-5
0.AB(V)
A.BC×10
-6
0.0A(V
A.BC×10
-7
0.00(V)
Fig. 9-5 REC-HOLD output (REC-HOLD with E-4)
0.01
0.1
1
10
1.0E-07 1.0E-06 1.0E-05 1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00
Output Voltage (V)
Pressure (Pa)