31
8.9. Selection Of DEGAS Function
Refer to P.39
Setting of DEGAS function
dGS: ANUAL MODE
dGS: AUTO MODE
DGS-①
The portion in the left figure blinks
after the segment indicator displays [
GS __
] for about 1 second. Each
time the [] key is pressed, the
position on the segment indicator
displays [
dGS _1
] [
dGS _2
]
cyclically.
To use DEGAS MANUAL, press the [
(ENTER)] key, with [
dGS _1
]
selected, to transfer control to the FIL-
status.
To use DEGAS AUTO, press the [
(ENTER)] key, with [
dGS _2
]
selected, to transfer control to the DGS-
status.
8.10. Setting of DEGAS Function
Refer to P.39
Set the numeric value of DEGAS auto OFF.
Settable numeric values are 01 to 99 minutes.
DGS-②
The portion in the left figure blinks.
Each time the [] key is pressed, the
numeric value of the blinking portion
changes. When a desired numeric
value is displayed, press the [
(ENTER)] key to transfer control to the
FIL- status.