43
9.6.4. Pseudo-LOG output
The entire measurable pressure range is outputted pseudo-logarithmically by an analog
voltage of 0 to 10 volts. The output voltage has hysteresis by the pressure that increased
and the pressure that decreased.
The table below gives the relationship between pressure and output voltage.
GI-M2 Pseudo-LOG OUTPUT
Table. 9-3 Measured Pressure Value Table (GI-M2)
Displayed pressure
value (Pa)
Displayed pressure
value (Pa)
×
×
×
×
×
×
×
×
×
×
×
-2
×
-7
×
×
×
×
Displayed pressure
value (Torr)
Displayed pressure
value (Torr)
×
-2
×
-6
×
×
×
×
×
-3
×
-8
×
×
×
×
×
-5
×
-9
×
×
Table. 9-4 Hysteresis by the pressure that increased and decreased
Displayed pressure
value (Pa)
Displayed pressure
value (Pa)
×
×
×
×
×
×
The measurement output voltage can be converted to a pressure value using the following
equation.
P:Pressure value (Pa)
V:Measured output voltage (V)
E:V from which decimal point is rounded down (V)
GI-M2:
P=10(V-E)×10
-(
8
-E)
(Pa)