5.CALIBRATION
39
• Setting of Scale Division (Omissible if no change is needed)
• Zero Calibration
1) Select setting mode 3.
2)
Select scale division.
Use and key to set the scale division,
then use key to validate the setting.
To return to the indicated value display, press key.
Setting of Scale Division
Scale Division(001 to 100)
1) Check that the sensor is unloaded,
then press key.
2) Press key.
If the indicated value following the display CALZE
is zero, the zero calibration is terminated.