Table 38 Error flags (read-only)
Register
number
Address Register
description
Data format Value
514 0202
hex
Barometer error
flags
16-bit unsigned
integer
0x0000 OK
0x0004 Pressure out of
range
0x0008 Pressure
measurement error
0x0010 Manufactory error
0x0020 Module type not
defined
0x0100 F meas
0x0400 Temperature meas
0x0800 Info 2nd bank
checksum
0x1000 Info 1st bank
checksum
0x2000 Checksum error of
current settings
0x4000 User setting bank
checksum
0x8000 Factory bank
setting checksum
Multiple device statuses can be present simultaneously. In those cases, the value
of the device status register is the sum of the applicable numbers, for example
0x14 if pressure is out of range (0x4) and a manufactory error (0x10) are present
simultaneously.