For the M20×1.5 cable gland ordered
together with the transmitter from Vaisala,
the cable diameter is 5.0 … 8.0 mm
(0.20 … 0.31 in). Tightening torque for the
cable gland is 8 Nm.
For the M20×1.5 cable gland with split
bushing, the cable diameter is 7 mm
(0.28 in).
Table 8 Analog output terminals
Terminal Function Notes
CH1 + Analog output channel 1 +
Max. wire size: 2.5 mm
2
(14 AWG)
CH1 - Analog output channel 1 -
CH2 + Analog output channel 2 +
CH2 - Analog output channel 2 -
CH3 + Analog output channel 3 +
CH3 - Analog output channel 3 -
CH4 + Analog output channel 4 +
CH4 - Analog output channel 4 -
3.3.5 Analog input terminals and lead-through
Before connecting wires or cables, make sure that the transmitter is powered o.
Analog input is not available in Indigo510 transmitters or transmitters that are powered with
Power over Ethernet (PoE).
Use the touchscreen or web interface to set the analog input and analog input powering on.
Indigo510, Indigo520 User Guide M212287EN-N
40