7.5 Enabling Modbus TCP/IP in web interface
You need Administrator user rights for this task.
1. Connect to the web interface.
2. Log in to the web interface as an administrator user.
3. Select Transmitter > Network.
4. Check that the network settings are set correctly.
5. Select Inputs and outputs > Modbus TCP/IP.
6. In Modbus o/on, set Modbus on.
7. Select Apply.
More information
‣
Modbus overview (page 94)
7.6
Disabling temperature compensation in web
interface
You need Administrator user rights for this task.
1. Connect to the web interface.
2. Log in to the web interface as an administrator user.
3. Select Transmitter > Compensation.
4. In Temperature compensation o/on, set compensation o.
5. Select Apply.
More information
‣
Temperature compensation (page 46)
Indigo510, Indigo520 User Guide M212287EN-N
92