Figure 37 Analog outputs test mode in web interface
1. Connect to the web interface.
2. Log in to the web interface as an administrator user.
3. Select Inputs and outputs > Analog outputs.
4. In Output mode, select the current or voltage range for the outputs.
5. In the Test mode tab, configure test mode settings.
a. In Test mode, set the test mode on for the analog output you want to test.
b. In Force output to, enter the current level that you want the transmitter to output or
choose Output error to use the transmitter's error state output level.
6. Select Apply.
7. Read the analog output and verify that the output is correct.
8. After verifying the output, go back to Inputs and outputs > Analog outputs > Test mode
and turn the test mode o.
9. Repeat step 5 … step 8 for all the analog output channels you want to test.