EasyManua.ls Logo

Vaisala Indigo500 Series - Verifying Cable Gland Tightness; Connecting Probes

Vaisala Indigo500 Series
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 10 Connection terminals for probe 1 and probe 2
Terminal Function Wire color in Vaisala
cables
Notes
RS-485- RS-485- White
Max. wire size: 2.5 mm
2
(14 AWG)
RS-485+ RS-485+ Black
24 VOUT - Power GND and
RS-485 common
Blue
24 VOUT + Positive supply voltage to
probe
Brown
3.3.8 Verifying tightness of cable glands
Adjustable wrench
To maintain enclosure tightness and provide strain relief to the cables, all cable glands must be
tightened. Unused cable glands must remain plugged.
1. Check every cable gland that is in use:
a. Pull on the cable to verify that the cable is securely held by the cable gland.
b. Tighten the cable gland if the cable moves.
2. Check that every unused cable gland is plugged and tightened.
3.4
 Connecting probes
The IP classification of probes is valid only when the probes are
connected to the probe connection cable.
CAUTION!
If you want to configure the probe settings, such as purge interval, you must do
that before connecting the probe to the transmitter. To configure the probe, you
can use the free Vaisala Insight PC software. For more information, see your
probe's user guide and visit www.vaisala.com/insight.
If your transmitter was delivered with preconfigured analog outputs, make sure
that you connect the probes accordingly. See the label inside the transmitter
enclosure for the correct order of the probes.
Indigo510, Indigo520 User Guide M212287EN-N
44

Table of Contents

Other manuals for Vaisala Indigo500 Series

Related product manuals