2.7.3 Relays
The Indigo520 transmitter provides 2 configurable relays that can be wired either as normally
closed or as normally open. Use the touchscreen or web interface to configure the relay
activation parameters.
Relays are not available in transmitters that are powered with Power over
Ethernet (PoE).
More information
‣
Configuring relays on touchscreen (page 80)
‣
Configuring relays in web interface (page 89)
2.8 Analog input
The Indigo520 transmitter provides 1 analog input channel with current input.
Current input mode: 4 ... 20 mA.
Using analog input is possible from software version 1.14.0 onward. To get the analog input
working, update your transmitter's software. No other changes to your transmitter are
required.
Analog input can be used as a source in analog output, relay and Modbus TCP/IP settings.
You can use Indigo520 to power your 2-wire current loop analog input device. If it is not
possible to power through Indigo520, or if you need 2 measurement probes, you can power
the analog input device with an external power supply.
Use the touchscreen or web interface to set the analog input and the analog input powering
on.
Only 1 measurement probe can be connected to the transmitter when you power
the analog input device through Indigo520 probe 2 connection terminal. The
measurement probe has to be connected to probe 1 connection terminal.
Analog input is not available in Indigo510 transmitters or transmitters that are
powered with Power over Ethernet (PoE).
More information
‣
Analog input terminals and lead-through (page 40)
‣
Configuring analog input on touchscreen (page 79)
‣
Configuring analog input in web interface (page 88)
‣
Updating transmitter software version (page 97)
Indigo510, Indigo520 User Guide M212287EN-N
20