9.4 Restoring factory default settings
You can restore all device settings to the factory default settings. This will take some minutes.
A factory reset deletes all current settings of the device. After the
factory reset, you need to reconfigure the settings, including outputs and relays.
When you connect to the web interface the next time, you will be prompted to
give the activation code and create new users.
CAUTION!
Allen key (4 mm), provided
A pointed object for pressing the button inside the transmitter.
1. Power o the transmitter.
2. Open the 2 hex screws on the cover of the transmitter using a 4-mm Allen key.
3. Open the transmitter cover.
4. Keep pressing the PRESS TO ACTIVATE WEB INTERFACE button while you power up the
transmitter. Keep pressing the button until the LED next to the button lights up.
ANALOG OUTPUTS
RELAY 1RELAY 2
+
+
+
+
+
+
+
+
+
CH1
CH2
CH3
CH4
RS-485
PROBE
24 VOUT
ANALOG
INPUT
NO COM NC NO COM NC
1
2
SERVICE PORT
PRESS TO ACTIVATE
WEB INTERFACE
ETHERNET
VS
+
The transmitter restarts.
If the transmitter asks you to calibrate the touchscreen display after the restart,
touch the requested crosshair positions on the touchscreen carefully and as
accurately as possible.
When the transmitter restarts after a factory reset, it will switch to using DHCP
and may get a new IP address. If needed, reconfigure the network settings using
the touchscreen. To continue using the web interface, reconnect using the
transmitter's new IP address.
Chapter 9 – Maintenance and troubleshooting