6. Configure the analog output channel settings:
a. In Output o/on, set the output channel o to edit settings.
b. In Source, select the source for the measurement parameter that controls the output
channel.
c. In Parameter, select the measurement parameter that controls the output channel.
d. In Scale low end and Scale high end, define the lower and upper ends of the
measurement scale for the chosen output parameter.
e. In Error output, define the output level indicating that the device is in error state.
f. In Low clipping limit, define the point (in mA below the output range) when the
output is clipped due to being out of range.
g. In High clipping limit, define the point (in mA or V above the output range) when the
output is clipped due to being out of range.
h. Select Apply.
7. Repeat step 5 … step 6 for all the analog output channels you want to use.
More information
‣
Analog output configuration example (page 78)
‣
Testing analog outputs in web interface (page 86)
7.2.1 Testing analog outputs in web interface
To test that the analog output of the transmitter is accurate, you can set the probe to output a
fixed level of current and view the reading to verify the output.
Indigo510, Indigo520 User Guide M212287EN-N
86