SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
148
panels, per discharge method.
Mean Depth
Average depth calculated from the total area
and width.
m or ft 3
Area
Total area calculated from all stations or
panels, per discharge method.
m
2
or ft
2
4
Mean SNR
Average SNR of all completed stations.
Require at least 3 stations.
dB n/a
Average temperature of all completed stations.
Average station velocity, X velocity component
Vel Min
Minimum station velocity, X velocity
component
m/s or ft/s 4
Vel Max
Maximum station velocity, X velocity
component
m/s or ft/s 4
Start staff gauge reading.
Uncertainty in discharge measurement.
Largest uncertainty in the discharge
uncertainty calculation.
% 2
GPS derived Latitude
DDD° MM' SS.SS
GPS derived Longitude
DDD° MM' SS.SS
Station Summary 8.3.5.4
The Station Summary from the data collection menu displays the completed stations in
table format. The station summary is a good reference when evaluating measurements
performed at a station against adjacent stations.
Station Summary screen consists of tabular view of
all completed stations,
a) Station Summary, Table 8:13.
b) To close the discharge summary report,
i). Press the Left Soft Key.
ii). The software will navigate to the Data
Collection Window.
Figure 8:44 - Station Summary
Properties associated with the variables and numerical calculations performed on the
raw data collected is defined in Table 8:13,
Table 8-13 - Station Summary