EasyManua.ls Logo

Xylem SonTek FlowTracker2 - 14.8.2.7 Diagnostics

Xylem SonTek FlowTracker2
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SonTek a Xylem brand
FlowTracker2 User’s Manual (January 2019)
272
Diagnostics 14.8.2.7
Figure 14:32. Layout of Diagnostics Tab
The Diagnostics tab (Figure 14:32) shows data pertaining to the quality of the
measurement. The following plots are available:
SNRthe mean SNR at the ADV sample volume over all samples and all beams
at each station. In the case of multiple point measurements per station, a value
will appear for each point measurements at a given station location. In the case
of multiple point measurements per station, a value will appear for each point
measurements at a given station location.
Spikethe number of spikes that were detected over all samples at each station.
Spikes are removed from the velocity measurement. In the case of multiple point
measurements per station, a value will appear for each point measurements at a
given station location.
Velocity standard errorthe velocity standard error calculated over all samples
at each station. In the case of multiple point measurements per station, a value
will appear for each point measurements at a given station location.
Samplesthe number of samples collected per station

Table of Contents