SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
124
The station delete process is not reversible and the user must ensure that the correct
station is selected. Selected station will be highlighted with a yellow background.
Delete Station is performed from the Data
Collection Window. Up to ten completed stations
will be displayed at any time,
a) To select a station,
i). Use the left or right arrow key to scroll through
the completed stations,
b) To DELETE a station,
i). Press the backspace key.
ii). The software will navigate to the Confirmation
Figure 8:9 - Select Station
Confirmation screen requests confirmation of
deleting the selected station by the FlowTracker2
handheld software.
a) To confirm station delete,
i). Press the Right Soft Key,
b) To CANCEL station delete,
i). Press the Left Soft Key,
c) The software will navigate to the Data Collection
Window.
Figure 8:10 - Confirm Delete
Edit Station 8.3.1.3
Editing the station parameters can be performed from the Data Collection Window by
selecting the station within the viewable window. The station editing process allows the
user to change all the parameters original configured when the station was created from
the “Add Station” screen.
Editing should not be performed on the Velocity Method after extended time has
passed since the initial point velocity measurements were performed. The flow
conditions could change during the lapsed time and additional point velocity
measurements will not be representative. It’s recommended that all the station
measurements be repeated if additional measurements are required.