EasyManua.ls Logo

Xylem SonTek FlowTracker2 - 14.8.1 Measurement Summary Pane

Xylem SonTek FlowTracker2
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SonTek a Xylem brand
FlowTracker2 User’s Manual (January 2019)
264
Figure 14:26. Layout of Data Viewing Window and Section Tab
Measurement Summary Pane 14.8.1
The Measurement Summary Pane values are summarized in Table 14:3.
Table 14-3. Measurement Summary Pane
Category
Value
Description
File Information
File name
File name
Start date and time
Start date and time of
measurement
Calculations engine
Identifies which calculation
algorithm was used: Original
FlowTracker (FlowTracker1) or
FlowTracker2
Data collection mode
Discharge or General Mode
System Information
Sensor type
Top setting, Universal, or Ice
(not available with Original
FlowTracker)
Handheld serial number
Handheld device serial number
Probe serial number
Probe serial number (not
available with Original
FlowTracker)
Probe firmware
Probe firmware
Handheld software
Handheld software version
Discharge Summary
# Stations
Number of stations measured
Mean depth
Mean depth over all stations
Mean velocity
Mean velocity over all velocity
measurements
Mean SNR
Mean Signal to Noise Ratio over
all velocity measurements
Mean temp
Mean temperature over all
measurements
Avg interval
Probe averaging interval per
sample (in seconds)
Total width
Total transect width
Total area
Total transect area
Total discharge
Total discharge
Discharge
Uncertainty
Category
ISO or IVE
Accuracy
Accuracy (%)
Depth
Uncertainty on depth (%)
Velocity
Uncertainty on velocity (%)
Width
Uncertainty on width (%)
Method
Uncertainty on method used (%)
# Stations
Uncertainty on number of
stations (%)

Table of Contents