Figure 15:2 – Upload of License File for Lab-ADV
15.3. Lab-ADV Configuration
The Lab-ADV Configuration settings specified in Table 15:1 allows the user to define the
data collection process within Communication, System, Environmental and Data
Management categories. The configuration settings are application dependent and will
vary depending on the specific measurement requirements.
Table 15-1. Configuration Settings
Category Value Description
Communication COM Port
COM port assigned to FlowTracker2 probe
assembly. It is recommended that the
Windows Device Manager be used to
determine assigned COM port.
Sampling
Sampling Rate
Sampling rate selected for measurement.
The Lab-ADV has four sampling rates: 1,
2, 5, and 10 Hz.
Velocity Range
Optimizes ADV operation for expected
maximum velocity conditions. There are
six velocity range settings: +/- 10, 20, 50,
100, 200 cm/s and Auto. Auto velocity
range is the most robust setting under
variable or unknown conditions. Auto is
not available at the10 Hz sample rate.