FlowTracker2 User’s Manual (January 2019)
171
Variable
Description
Units
Decimal
Temp
Mean temperature during measurement
°C or °F
2
Depth
Mean water depth from water surface during
point measurement
m or ft 3
Pressure
Mean water pressure during point
measurement
dbar 3
Review Station Measurement 9.3.3.3
Review Station measurement is a tabular report that is available from the data collection
window at the completion of point velocity measurement of a station.
Review Station screen consists of tabular summary
of measurements performed at the station,
a) Review Station, Table 9:7.
b) To access review station report,
i). Select the station in the data collection window,
ii). Press the enter key.
c) To EXIT review station report,
i). Press the Left Soft Key.
Figure 9:25 - Review Station
Properties associated with the variables and numerical calculations performed on the
raw data collected at the station is defined in
Table 9:7,
Table 9-7 - Review Station
Variable
Description
Units
Decimal
Loc X
Location X coordinate
m or ft
3
Loc Y
Location Y coordinate
m or ft
3
Depth
Measured water depth
m or ft
3
Meas. Depth
Velocity measurement depth
m or ft
3
Comment
Comment supplied before the measurement
n/a
n/a
Warning
Warning supplied based on quality control
criteria
n/a n/a
Samples
Total number of sample received
number
n/a
Spikes
Number of spikes removed from mean velocity
number
n/a
Vel. X
Mean X velocity component of all samples.
Velocity is despiked and mounting correction
applied (if applicable).
m/s or ft/s 4
Vel. Y
Mean Y velocity component of all samples.
Velocity is despiked and mounting correction
applied (if applicable).
m/s or ft/s 4
Vel. Z
Mean Z velocity component of all samples.
Velocity is despiked and mounting correction
applied (if applicable).
m/s or ft/s 4
σV. X
Standard error of X velocity component.
m/s or ft/s
4