SonTek a Xylem brand
FlowTracker2 User’s Manual (January 2019)
98
Warning Reporting QC Criteria Description Suggested Action
Fractional
Depth > 1
Point
Velocity
None
Fractional Depth < 1. Verify ratio
between measurement depth and
total depth and this should not >1
Verify fractional depth
value.
Re-enter if needed.
Timing of Warning Messages 6.2.2
The quality control warning messages are supplied at certain stages of the measurement
process for both the Discharge and General Modes. The timing of the quality control
warning messages are supplied in Table 6:5. D” indicates Discharge mode and “G”
indicates General mode.
Table 6-5 - Timing of Quality Control Warning Messages
Warning Messages
Station
Setup
Pre
Measurement
Post
Measurement
Post Station
Post
Section
Low SNR
D,G
D,G
Approach Low SNR
D,G
D,G
Beam SNRs Not
Similar
D,G
D,G
Large SNR Variation
D,G
D,G
SNR Threshold
Variation
D,G
D,G
Standard Error > QC
D,G
D,G
Boundary Interference
D,G
High % Spikes
D,G
D,G
Velocity Angle > QC
D
D
Rod Angle > QC
D,G
D,G
High Stn % Discharge
D
D
Water Depth > QC
D
D
Stn Spacing > QC
D
Station Order
D
Location Outside Edge
D
Fractional Depth > 1
D,G
D,G
6.3. Beam Check
Beam Check operates by sending a pulse of sound into the water, and then plotting the
signal to noise ratio of the return signal versus range for each of the FlowTracker2
receivers. This information can be evaluated to determine the effective measurement
range, to look for interference from boundaries/structures, to survey a deployment site,
or to observe the quality of the returned signal.