EasyManua.ls Logo

Xylem SonTek FlowTracker2 - 3.2.3 File Naming; 3.2.4 Folder Naming

Xylem SonTek FlowTracker2
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SonTek a Xylem brand
FlowTracker2 User’s Manual (January 2019)
34
i). Top-Setting (Instrument referenced to water surface)
ii). Universal (Instrument referenced to bottom)
iii). Ice (Instrument referenced to water surface)
iv). Unknown
b) The graphic representation of the Wading Rods supported and the fractional
depth indicators are defined in Table 3:3,
Table 3-3 - Wading Rod Graphics
Wading Rod
Graphic
Wading Rod Setting
Example
Reference
Top Setting
= 󰇩
1

×

0.4
󰇪
f
depth
Fractional Depth
w
depth
Water Depth
w
depth
= 0.783m
f
depth
= 0.8
Rod = 0.391
Water
Surface
Universal
=

󰇡

×

󰇢
f
depth
Fractional Depth
w
depth
Water Depth
w
depth
= 0.783m
f
depth
= 0.6
Rod = 0.313
Channel
Bed
Ice
=

󰇣
󰇡

depth ice
󰇢
×
󰇡
1

󰇢
󰇤
f
depth
Fractional Depth
w
depth
Water Depth
I
thickness
= Ice Thickness
w
depth ice
= Water Surface
to Bottom of Ice
w
depth
= 0.783m
f
depth
= 0.5
I
thickness
= 0.2m
w
depth ice
= 0.1m
Rod = 0.442
Water
Surface
c) To change the Wading Rod type,
i). Use the left or right arrow key to scroll through the list of options,
ii). Press the Left Soft Key to return to Configuration for the software to accept the
change.
File Naming 3.2.3
The File Naming function enables the operator to assign a unique name for each
measurement file. The file naming process in FlowTracker2 ensures that existing
measurement files are not overwritten.
a) The file naming conventions that are supported in FlowTracker2 software are
listed in Table 3:4,

Table of Contents