SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
170
f) Velocity Angle,
g) Probe Depth,
h) Battery,
i) Pressure.
j) To select a variable,
i). Use the up or down arrow key to view a
variable,
k) To accept the velocity measurement,
i). Press the Right Soft Key.
l) To REDO velocity measurement,
i). Press the Left Soft Key.
The conventions and graphic display features used for displaying Velocity and SNR raw
data are defined under Raw Data Display.
Properties associated with the variable output from the numerical calculations
performed on raw data collected is defined in Table 9:6,
Table 9-6 - Review Point Measurement
Warning
Warning supplied based on quality control
criteria
n/a n/a
Total number of sample received
Number of spikes removed from mean velocity
Vel. X
Mean X velocity component of all samples.
Velocity is despiked and mounting correction
m/s or ft/s 4
Vel. Y
Mean Y velocity component of all samples.
Velocity is despiked and mounting correction
m/s or ft/s 4
Vel. Z
Mean Z velocity component of all samples.
Velocity is despiked and mounting correction
m/s or ft/s 4
Standard error of X velocity component.
Standard error of Y velocity component.
Standard error of Z velocity component.
Signal to noise ratio of Beam 1
Signal to noise ratio of Beam 2
Signal to noise ratio of Beam 3
Flow angle relative to X direction
Tilt
Mean wading rod angle during point velocity
measurement
Degree 2