SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
184
The communication window displays the following
instructions to enable connection between the
handheld and FlowTracker2 desktop software,
a) Open FlowTracker2 desktop software,
b) Select “Device” menu option,
c) Is USB cable connected or Bluetooth enabled,
d) Select communication medium.
e) To CLOSE the communication function,
i). Press the Left Soft Key.
Figure 11:2 - Communication
Connect FlowTracker2 desktop software with
handheld using either USB or Bluetooth
communication.
a) Connect to FlowTracker2 handheld Device,
i). Select “Device” menu option on main menu
banner,
ii). Connection window will appear on software
window with USB and Bluetooth
communication options.
b) Select communication medium,
i). USB - Make sure USB cable is connected
before USB option is selected,
ii). Bluetooth – Select handheld serial number from
dropdown list and select the “Connect”
Figure 11:3 - Device
Communication Window
If the handheld serial number is not listed, select the content in the box using mouse
and start typing the serial number. When the serial number is entered, select “Connect”
function. The desktop software will connect to the handheld using bluetooth and the
serial number will be stored for future use.
Perform Upgrade of handheld software and or ADV
firmware by selecting the upgrade function on the
FlowTracker2 desktop software.
a) Perform Upgrade of FlowTracker2 device,
i). Select “Upgrade” menu option,
ii). Open file window will appear with the available
firmware file.
iii). Select the firmware file within the “Open File”
window and select “Open” function.
Figure 11:4 - Upgrade Function