EasyManua.ls Logo

Xylem SonTek FlowTracker2 - Page 290

Xylem SonTek FlowTracker2
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SonTek a Xylem brand
FlowTracker2 User’s Manual (January 2019)
289
Category Value Description
Logging Duration Per
File (min)
Defines a file logging limit based on a time
duration.
Logging Interval Start /
End Time Alignment
The Logging Interval Start / End Time
Alignment works in combination with
“Logging Duration Per File (min)”.
If your logging duration per file is specified
as 10 min and you start your logging at
17:03:25:
- Without logging interval alignment
the next file will be started at
17:13:25, the third one at 17:23:25
and so on
- With logging interval alignment the
next file will be started at 17:20:00,
the third one at 17:30:00 and so on
Table 15-2: Velocity Range Settings
Velocity
Range
Absolute max
velocity
in arbitrary
direction
in ideal
conditions
Absolute max
velocity
in X direction
in ideal
conditions
Recommended
max velocity
in arbitrary
direction
in ideal
conditions
Recommended
max velocity
in X direction
in ideal
conditions
Auto
± 112 cm/s
± 448 cm/s
± 60 cm/s
± 200 cm/s
± 10 cm/s
± 10 cm/s
± 40 cm/s
± 3 cm/s
± 10 cm/s
± 20 cm/s
± 14 cm/s
± 52 cm/s
± 7 cm/s
± 20 cm/s
± 50 cm/s
± 27 cm/s
± 108 cm/s
± 15 cm/s
± 50 cm/s
± 100 cm/s
± 55 cm/s
± 220 cm/s
± 30 cm/s
± 100 cm/s
± 200 cm/s
± 112 cm/s
± 448 cm/s
± 60 cm/s
± 200 cm/s
Ideal conditions consist of:
- ideal signal conditions (high SNR, low noise)
- Ideal boundary conditions (no boundaries nearby)
- Ideal temperature and sound speed (18°C and 1475 m/s)
- For X direction velocity uniform flow in X direction
2D/3D FT2 probe would use the same recommendation with Z velocity maximum
same as the arbitrary direction.

Table of Contents