9-7
IM 701510-01E
Performing a GO/NO-GO Action
9
Points to note relating to zone determination
•Trigger mode is automatically switched to single short mode/single mode irrespective of
the selected trigger mode during GO/NO-GO determination. However, determination will
be repeated if the determination sequence has been set to “CONTINUE”. Even in case the
GO/NO-GO mode (at the GO/NO-GO top menu) is changed to OFF again, the setting at
the trigger mode selection will not become valid again.
It is not possible to perform zone determination while both pre-zoom waveform and its
expanded waveform are displayed at the same time or while an FFT or X-Y waveform is
displayed. Neither is it possible to a display zone waveform. To perform GO/NO-GO
determination using a zone waveform, make sure that the expanded waveform alone is
displayed before “Mode” is set to “ON” (at the GO/NO-GO top menu).
The following operations are not possible after having pressed the “EXEC” soft key. In
order to perform these operations, it is necessary to set “Mode” (at the GO/NO-GO top
menu) to “OFF”, after having pressed the “ABORT” soft key.
• Initialization of the settings Automated measurements
Auto set-up • Changing of expansion settings
• Cursor measurements • Display of an X-Y waveform
Action on trigger
•A zone waveform is not displayed if the following conditions are not satisfied after the
zone waveform has been created.
• “Mode” at the GO NO-GO top menu : ON
• “GO/NO-GO Type” : ZONE
• “Mode” for selected trace : ON
• Display for selected trace : ON (corresponding LED is lit)
During the time in which an action is being executed after GO/NO-GO determination, no
waveform data will be acquired.
Note
The zone waveform generated can be retained by saving the set-up data on a floppy disk. However,
“Mode” (at the GO NO-GO top menu) must be set to “ON” and “GO/NO-GO Type” to “ZONE”. For a
description of how to save set-up data on a floppy disk, refer to 12.4 “Saving and Loading Set-up Data”
(page 12-7).
9.1 Judging using a Waveform Zone