5-4
IM 701510-01E
5.3 Setting the Probe Attenuation
1. Press the INPUT key corresponding to the channel for which the
probe attenuation is to be selected. The input setting menu will be
displayed.
2. Press the “Probe” soft key to display the probe selection menu.
3. Press the soft key corresponding to the desired attenuation.
Explanation
Probe attenuation
Probe attenuation can be selected from “1:1”, “10:1”, “100:1” and “1000:1”. The default is
“10:1”.
Set to “10:1” when using the probe supplied with the instrument.
Set to “1:1” when using a BNC cable.
Note
The selected probe attenuation will remain unchanged even if auto set-up is performed.
When the probe attenuation is changed, the offset voltage also changes accordingly.
Keys and Procedure
INPUT key
Soft keys