9-12
IM 701510-01E
Points to note relating to parameter determination
•Trigger mode is automatically switched to single mode irrespective of the selected trigger
mode during GO/NO-GO determination. However, determination will be repeated if the
determination sequence has been set to “CONTINUE”. Even in case the GO/NO-GO
mode (at the GO/NO-GO top menu) is changed to OFF again, the setting at the trigger
mode selection will not become valid again.
While both pre-zoom waveform and its expanded waveform are displayed at the same
time, the expanded waveform will become the object of determination.
The following operations are not possible after having pressed the “EXEC” soft key. In
order to perform these operations, it is necessary to set “Mode” (at the GO/NO-GO top
menu) to “OFF”, after having pressed the “ABORT” soft key.
• Initialization of the settings Automated measurements
Auto set-up • Changing of expansion settings
• Cursor measurements • Display of an X-Y waveform
• Action-on trigger
During an action is being executed after GO/NO-GO determination, no waveform data
will be acquired.
In case the linear scaling function has been set to ON, you can enter linear scaled values
at the GO/NO-GO parameter setting menu.
Only waveforms displayed on screen are available for judgment. Acquisition-memory
data is not available.
9.2 Judging using Measured Values of Waveform Parameters