EasyManuals Logo

YOKOGAWA DLM3024 User Manual

YOKOGAWA DLM3024
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
5-2
IM DLM3054-01EN
Contents of Emails That the Instrument Sends (Go/Nogo AND, Go/Nogo OR)
<Subject>: The subject attached to the email. For example: “GoNogo Triggered Report (Nth Nogo result)”
[Comment]: Comment
[Setup Information]: Reference conditions (contents of condition numbers 1 to 4), Logic (AND/OR), Stop
Nogo/Action Count ((the number of no-go results/the number of actions)
[Trigger Date and Time]: The time of trigger occurrence
[Nogo/Exec Count]: The number Nogo results/the number of judgments performed
[Nogo Factor]: Information, including measured values, about the conditions that returned no-go results*
* Measured values are only sent for GO/NO-GO determination based on waveform parameters.
Example
[Comment] Test
[Setup Information]
Select1: Rect(C1,Main) (Left:-3.0000E+00,Right:-2.5000E+00,Upper: 500.00E-03,Lower:-
500.00E-03,Condition:In)
Select2: Wave(C1,Main) (Range1:-5.0000E+00,Range2: 5.0000E+00,Condition:In)
Select3: Polygon(C1,Main) (Condition:In)
Select4: Measure(Max(C1)) (Upper: 0.0000E+00,Lower: 0.0000E+00,Condition:In)
Logic: OR
Stop Nogo/Action Count: 1/Inf
[Trigger Date and Time] 2018/06/01 21:18:51
[Nogo/Exec Count] 1/1
[Nogo Factor]
Select1: Rect(C1,Main)
Select2: Wave(C1,Main)
Number of Actions (Action Count/Nogo Count)
When the number of waveform acquisitions reaches the specified action count or when the number of no-go
results reaches the specified Nogo Count, the instrument stops waveform acquisition.
Action Count
Set the number of waveform acquisitions.
1 to 1000000: The instrument stops the action when it acquires the specified number of waveforms.
Infinite (0): The instrument repeats the action until waveform acquisition is stopped using the Abort soft key or
the RUN/STOP key.
Nogo Count
Set the number of no-go results.
1 to 1000: When the specified number of no-go results is reached, the instrument stops GO/NO-GO
determination.
Infinite (0): The instrument repeats GO/NO-GO determination until waveform acquisition is stopped using the
Abort soft key or the RUN/STOP key.
5 Executing Actions

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM3024 and is the answer not in the manual?

YOKOGAWA DLM3024 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM3024
CategoryTest Equipment
LanguageEnglish

Related product manuals