Maintenance and Diagnostic Procedures Section 5
5-20 Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
5. Inspect the heater blocks for corrosion or foreign matter
and clean as appropriate.
Figure 5.3: Inspect Heater Block for Corrosion
6. Inspect the bottom of the heated basins for corrosion or
foreign matter and clean or replace as appropriate.
7. Inspect the heated basin coating for breaches or bubbling
and replace the basin if necessary.
8. Fill each heated basin with 500 mL of room temperature
water and place in the heated basin locations on the
Abbott VP 2000 Processor. Make sure the basins are firmly
seated in the heater blocks.
9. Close the heater lid and check to see that each sensor
probe is completely immersed in its basin water.