Section 4 Instrument Installation
Abbott VP 2000™ Processor Service Manual 4-13
30-608308/R1—September 2008
Allow access to electrical circuit with voltage appropriate
to the model. This circuit should be separate, with its own
fuse or circuit breaker, shared with no other instrument.
Allow access to a water supply (DI recommended) and
drain facilities lower than the output of the unit since
water drains by gravity.
Leveling the Unit
Level the unit front to back and left to right by turning the
adjustable feet with a 3/8” open-end wrench.
CAUTION: Non-level operation could cause improper
reagent and water levels, resulting in spill over.