Maintenance and Diagnostic Procedures Section 5
5-38 Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Mainframe Test and Setup Checklist
Table 5.3: Mainframe Test and Setup Checklist
Abbott VP 2000 Processor Mainframe Test and Setup Checklist
MSI P/N SERIAL NO.
PROM VERSION NO. INPUT VOLTAGE
ELECTRICAL TEST PASS FAIL REPAIR RETEST TEST IAW
ON/OFF Switch  4022-18594, 3.1
Home Function

4022-18594, 3.1
PROM Version
 4022-18594, 3.2
Self Test
 4022-18594, 3.3
Power Interrupt Alarm
 4022-18594, 3.4
Stored Program Test
 4022-18594, 3.7
Heater Function
 4022-18594, 3.8
Interlock Switch Test
 4022-18594, 3.9
Sensor Voltage Test
 4022-18594, 3.10
Y Motor Torque Test
 4022-18594, 3.11
Arm Fail Safe
 4022-18594, 3.12
Printer Output
 4022-18594, 3.13
Computer Interface
 4022-18594, 3.14
FLUID CHECK PASS FAIL REPAIR RETEST TEST IAW
Flow Rate Adjust  4022-18594, 4.1
Flow Action

4022-18594, 4.1
Recirculate
 4022-18594, 4.2
Drain Action
 4022-18594, 4.2
VISUAL INSPECTION PASS FAIL REPAIR RETEST TEST IAW
X Motor Mounting Screw  4022-18594, 6.1
Y Table Wire

4022-18594, 6.2
TESTED BY: DATE:
QC APPROVAL DATE