Maintenance and Diagnostic Procedures Section 5
5-32 Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
22. The temperature should rise gradually to 80
o
C 5
o
C over
the course of the protocol and remain within 5
o
C of
80
o
C until the end of the protocol.
23. If the temperature fails to reach the target 80
o
C 5
o
C
check for restricted air flow or malfunctioning Air Heater
Element, correct as necessary, and retest.