Section 6 Maintenance Diagnostic Faults
Abbott VP 2000™ Processor Service Manual 6-17
30-608308/R1—September 2008
Figure 6.1: No X Movement
NO X MOVEMENT
INSPECT X DRIVE BELT
1
PASS FAIL
REPAIR OR REPLACE
INSPECT CARRIAGE
MOVEMENT
2
PASS FAIL
INSPECT X MOTOR
PULLEY
PASS FAIL
3
INSPECT IDLER PULLEY
PASS FAIL
4
INSPECT CONNECTORS
FAILPASS
5
REPLACE X MOTOR REPLACE PULLEY ASSY
TEST X SENSORS
PASS FAIL
TEST X MOTOR
FAILPASS
6
7
REALIGN & PLUG IN
REPLACE X SENSOR
BOARD
REPLACE X MOTOR
COMPUTER BOARD
8
TEST BY SUBSTITUTION
POWER SUPPLY
BOARD
TEST BY SUBSTITUTION
8
TEST CONTROL
MODULE
TEST BY SUBSTITUTION
8

Table of Contents