Troubleshooting Section 6
6-42 Abbott VP 2000™ Processor Service Manual
30-608308/R1—September 2008
Figure 6.16: Fluid Flow Rate Incorrect
FLUID FLOW RATE INCORRECT
MEASURE & ADJUST REGULATOR
FLOW
FAIL
1
REPLACE FLUID-IN
CONTROL ASSEMBLY