R3131 Spectrum Analyzer Operation Manual
3.3 Menu Function Descriptions
3-39
(2) When using the delta marker:
Delta → CF Sets the frequency difference between the delta and normal
marker as the center frequency.
Delta → CF step Sets the frequency difference between the delta and normal
marker as the step size for the center frequency.
Delta → Span Sets the frequency difference between the delta and normal
marker as the frequency span.
Delta → MKR Step Sets the frequency difference between the delta and normal
marker as the step size of the marker.
MKR Trace Selects one of the two traces so that the marker is available when
traces A and B are displayed simultaneously.
1/2, more Displays the Mkr → (2) menu.
MKR → CF Sets the active marker frequency as the reference level.
MKR → Ref Sets the active marker level as the reference level.
Peak → CF Moves the marker to the maximum peak within the search range,
and sets the marker frequency as the center frequency.
Peak → Ref Moves the marker to the maximum peak within the search range,
and sets the marker level as the reference level.
MKR → CF step Sets the active marker frequency as the step size of the center
frequency.
MKR → MKR Step Sets the active marker frequency as the step size of the marker.
2/2, more Returns to the Mkr → (1) menu.
Apr 10/98