R3131 Spectrum Analyzer Operation Manual
3.3 Menu Function Descriptions
3-38
3.3.15 MKR → Key (Marker →)
This section describes the menu displayed when the Mkr
→
key is pressed. This menu allows you to use
the active marker data (such as frequency and level) as the data for some other function. Item(s) in the Mkr
→
(1) menu may be different depending on the marker you use (either normal or delta marker).
(1) When using the normal marker:
MKR
→
CF Sets the active marker frequency as the center frequency.
MKR
→
Ref Sets the active marker level as the reference level.
Peak
→
CF Moves the marker to the maximum peak within the search range,
and sets the marker frequency as the center frequency.
Peak
→
Ref Moves the marker to the maximum peak within the search range,
and sets the marker level as the reference level.
MKR Trace Selects one of the two traces so that the marker is available when
traces A and B are displayed simultaneously.
1/2, more Displays the Mkr
→
(2) menu.
MKR
→
CF step Sets the marker frequency as the center frequency for the step
size.
MKR
→
MKR Step Sets the marker frequency as the step size of the marker.
2/2, more Returns to the Mkr
→
(1) menu.
Apr 10/98