EasyManua.ls Logo

ADVANTEST R3131

Default Icon
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R3131 Spectrum Analyzer Operation Manual
3.3 Menu Function Descriptions
3-37
Fixed MKR ON/OFF Toggles the Fixed Marker function on or off.
ON: Stores the frequency and level of the displayed delta
marker and fixes the marker at the current physical
position on the screen.
OFF: Turns off the Fixed Marker function.
MKR Step AUTO/MNL Toggles the step size function on or off. This function allows you
to control the movement of the marker using the step key.
AUTO: Automatically sets the step size to 1/10 of the span.
MNL: Allows you to manually set the step size to any value.
Marker Couple ON/OFF Toggles the Marker Couple function on or off.
ON: Specifies the peak search range used within the
measuring window.
OFF: Turns off the Marker Couple function.
2/2, more Returns to the Marker (1) menu.
Apr 10/98

Table of Contents

Related product manuals